7: TSP command reference Series 2600B System SourceMeter® Instrument
7-346 2600BS-901-01 Rev. B / May 2013
Example
trigger.timer[3].delaylist = {50e-6, 100e-6, 150e-6}
DelayList = trigger.timer[3].delaylist
for x = 1, table.getn(DelayList) do
print(DelayList[x])
end
Set a delay list on trigger timer 3
with three delays (50 µs, 100 µs,
and 150 µs).
Read the delay list on trigger
timer 3.
Output (assuming the delay list was
set to 50 µs, 100 µs, and 150 µs):
5.000000000e-05
1.000000000e-04
Also see
trigger.timer[N].reset() (on page 7-348)
trigger.timer[N].EVENT_ID
This constant specifies the trigger timer event number.
Type TSP-Link accessible Affected by Where saved Default value
Constant Yes
Usage
eventID = trigger.timer[N].EVENT_ID
Trigger timer number (1 to 8)
Details
This constant is an identification number that identifies events generated by this timer.
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to events
from this timer.
Example
trigger.timer[1].stimulus = tsplink.trigger[2].EVENT_ID
Sets the trigger stimulus of
trigger timer 1 to the TSP-
Link trigger 2 event.
Also see
None
trigger.timer[N].overrun
This attribute indicates if an event was ignored because of the event detector state.
Type TSP-Link accessible Affected by Where saved Default value
Recall setup
Trigger timer N clear