System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. B / May 2013 7-265
As an example, to set bit B1 of the measurement event reading overflow summary enable register, set
status.measurement.reading_overflow.enable =
status.measurement.reading_overflow.SMUA.
In addition to the above constants, measurementRegister can be set to the numeric equivalent of the bit to
set. To set more than one bit of the register, set measurementRegister to the sum of their decimal weights.
For example, to set bits B1 and B2, set measurementRegister to 6 (which is the sum of 2 + 4).
Example
status.measurement.reading_overflow.enable =
status.measurement.reading_overflow.SMUA
Sets the SMU A bit of the
measurement reading overflow
summary enable register using a
Also see
Measurement event registers (on page E-8)
status.measurement.voltage_limit.*
This attribute contains the measurement event voltage limit summary register set.
Type TSP-Link accessible Affected by Where saved Default value
.enable (RW) Yes Status reset Not saved 0
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B: 6
Usage
measurementRegister = status.measurement.voltage_limit.condition
measurementRegister = status.measurement.voltage_limit.enable
measurementRegister = status.measurement.voltage_limit.event
measurementRegister = status.measurement.voltage_limit.ntr
measurementRegister = status.measurement.voltage_limit.ptr
status.measurement.voltage_limit.enable = measurementRegister
status.measurement.voltage_limit.ntr = measurementRegister
status.measurement.voltage_limit.ptr = measurementRegister
The status of the measurement voltage limit summary register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other decimal values
indicate various bit settings
Details
These attributes read or write to the measurement event voltage limit summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is bit B0, and the most significant bit is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to Status register set contents (on
page E-1) and Enable and transition registers (on page E-19). The individual bits of this register are defined in
the following table.