Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-329
Example 2
-- decimal 18 = binary 0000 0000 0001 0010
enableRegister = 18
status.system5.enable = enableRegister
Uses a decimal value to set bits B1 and
B4 of the system summary 5 enable
register.
Also see
status.system4.* (on page 7-325)
System summary and standard event registers (on page E-7)
SweepILinMeasureV()
This KISweep factory script function performs a linear current sweep with voltage measured at every step (point).
Type TSP-Link accessible
Affected by Where saved Default value
Function No
Usage
SweepILinMeasureV(smu, starti, stopi, stime, points)
smu
System SourceMeter
®
instrument channel (for example, smua refers to SMU
channel A)
starti
Sweep start current in amperes
stopi
Sweep stop current in amperes
stime
Settling time in seconds; occurs after stepping the source and before performing a
measurement
points
Number of sweep points (must be ≥2)
Details
Data for voltage measurements, current source values, and timestamps are stored in smuX.nvbuffer1.
If all parameters are omitted when this function is called, this function is executed with the parameters set to the
default values.
Performs a linear current sweep with voltage measured at every step (point):
1. Sets the smu to output starti amperes, allows the source to settle for stime seconds, and then
performs a voltage measurement.
2. Sets the smu to output the next amperes step, allows the source to settle for stime seconds, and then
performs a voltage measurement.
3. Repeats the above sequence until the voltage is measured on the stopi amperes step.
The linear step size is automatically calculated as follows:
step = (stopi – starti) / (points – 1)
Figure 136: SweepILinMeasureV()

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals