Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-260 2600BS-901-01 Rev. B / May 2013
Bit Value Description
B0
Not used
Not applicable.
B1
status.measurement.current_limit.SMUA
Set bit indicates that the SMU A current limit
was exceeded.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2
status.measurement.current_limit.SMUB
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set
bit indicates that the SMU B current limit was
exceeded.
Bit B2 decimal value: 4
Binary value: 0000 0100
B3-B15
Not used
Not applicable.
As an example, to set bit B1 of the measurement event current limit summary enable register, set
status.measurement.current_limit.enable = status.measurement.current_limit.SMUA.
In addition to the above constants, measurementRegister can be set to the decimal equivalent of the bit to
set. To set more than one bit of the register, set measurementRegister to the sum of their decimal weights.
For example, to set bits B1 and B2, set measurementRegister to 6 (which is the sum of 2 + 4).
Example
status.measurement.current_limit.enable =
status.measurement.current_limit.SMUA
Sets the SMUA bit of the
Measurement Event Current Limit
Summary Enable Register.
Also see
Measurement event registers (on page E-8)
status.measurement.instrument.smuX.* (on page 7-262)
status.measurement.instrument.*
This attribute contains the registers of the measurement event instrument summary register set.
Type TSP-Link accessible Affected by Where saved Default value
Attribute - - - - - - - -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B:
6 (All bits set)
Usage
measurementRegister = status.measurement.instrument.condition
measurementRegister = status.measurement.instrument.enable
measurementRegister = status.measurement.instrument.event
measurementRegister = status.measurement.instrument.ntr
measurementRegister = status.measurement.instrument.ptr
status.measurement.instrument.enable = measurementRegister
status.measurement.instrument.ntr = measurementRegister
status.measurement.instrument.ptr = measurementRegister

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals