System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. B / May 2013 7-239
smuX.trigger.endsweep.action
This attribute sets the action of the source at the end of a sweep.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Usage
action = smuX.trigger.endsweep.action
smuX.trigger.endsweep.action = action
The source action at the end of a sweep; set to one of the following values:
0 or smuX.SOURCE_IDLE: Sets the source level to the programmed (idle) level at
the end of the sweep
X
.SOURCE_HOLD: Sets the source level to stay at the level of the last step
Source-measure unit (SMU) channel (for example,
smua.trigger.endsweep.action applies to SMU channel A)
Details
Use this attribute to configure the source action at the end of the sweep. The SMU can be programmed to return
to the idle source level or hold the last value of the sweep.
Example
smua.trigger.endsweep.action =
smua.SOURCE_IDLE
Sets SMU channel A to return the source
back to the idle source level at the end of a
Also see
Triggering (on page 3-32)
smuX.trigger.IDLE_EVENT_ID
This constant contains the idle event number.
Type TSP-Link accessible Affected by Where saved Default value
Constant Yes
Usage
eventID = smuX.trigger.IDLE_EVENT_ID
Source-measure unit (SMU) channel (for example,
smua.trigger.IDLE_EVENT_ID applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to idle events
from this SMU.