Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-367
Example
tspnet.write(testdevice, "print([[hello]])")
print(tspnet.readavailable(testdevice))
tspnet.clear(testdevice)
print(tspnet.readavailable(testdevice))
Write data to a device, then print how much is
available.
Output:
6.00000e+00
Clear data and print how much data is
available again.
Output:
0.00000e+00
Also see
tspnet.connect() (on page 7-367)
tspnet.readavailable() (on page 7-371)
tspnet.write() (on page 7-376)
tspnet.connect()
This function establishes a network connection with another LAN instrument or device through the LAN interface.
Type TSP-Link accessible
Affected by Where saved Default value
Function No
Usage
connectionID = tspnet.connect(ipAddress)
connectionID = tspnet.connect(ipAddress, portNumber, initString)
connectionID
The connection ID to be used as a handle in all other
tspnet
function calls
ipAddress
IP address to which to connect in a string
portNumber
Port number (default 5025)
initString
Initialization string to send to
ipAddress
Details
This command connects a device to another device through the LAN interface. If the portNumber is 23, the
interface uses the Telnet protocol and sets appropriate termination characters to communicate with the device.
If a portNumber and initString are provided, it is assumed that the remote device is not TSP-enabled. The
Series 2600B does not perform any extra processing, prompt handling, error handling, or sending of commands.
In addition, the tspnet.tsp.* commands cannot be used on devices that are not TSP-enabled.
If neither a portNumber nor an initString is provided, the remote device is assumed to be a Keithley
Instruments TSP-enabled device. Depending on the state of the tspnet.tsp.abortonconnect attribute, the
Series 2600B sends an abort command to the remote device on connection.
The Series 2600B also enables TSP prompts on the remote device and error management. The Series 2600B
places remote errors from the TSP-enabled device in its own error queue and prefaces these errors with Remote
Error, followed by an error description.
Do not manually change either the prompt functionality (localnode.prompts) or show errors by changing
localnode.showerrors on the remote TSP-enabled device. If you do this, subsequent tspnet.tsp.*
commands using the connection may fail.
You can simultaneously connect to a maximum of 32 remote devices.

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals