Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-46 2600BS-901-01 Rev. B / May 2013
dataqueue.CAPACITY
This constant is the maximum number of entries that you can store in the data queue.
Type TSP-Link accessible
Affected by Where saved Default value
Constant
Yes
Usage
count = dataqueue.CAPACITY
count
The variable that is assigned the value of
dataqueue.CAPACITY
Details
This constant always returns the maximum number of entries that can be stored in the data queue.
Example
MaxCount = dataqueue.CAPACITY
while dataqueue.count < MaxCount do
dataqueue.add(1)
end
print("There are " .. dataqueue.count
.. " items in the data queue")
This example fills the data queue until it is full
and prints the number of items in the queue.
Output:
There are 128 items in the data
queue
Also see
dataqueue.add() (on page 7-45)
dataqueue.clear() (on page 7-46)
dataqueue.count (on page 7-47)
dataqueue.next() (on page 7-48)
dataqueue.clear()
This function clears the data queue.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
dataqueue.clear()
Details
This function forces all dataqueue.add() commands that are in progress to time out and deletes all data from
the data queue.

Table of Contents

Other manuals for Keithley 2601B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601B and is the answer not in the manual?

Keithley 2601B Specifications

General IconGeneral
BrandKeithley
Model2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals