The M4 TORNADO system
32
User Manual
Parameter M4 TORNADO
Sample positioning
Stage
Traversing range
Mapping range
Max. speed
motorized X-Y-Z-stage
200 mm x 160 mm x 120 mm
190 mm x 160 mm
100 mm/s
Sample view
Camera 1, 2
Magnification
Camera 3
PC connection
coloured camera system with microscope
10x and 100x sample view of approx. 1 mm² and 1 cm²
door camera for chamber overview
USB
Operating conditions
Optimum temp. range
Working temp. range
Rel. humidity
Compartment air
22 °C – 28 °C
18 °C – 32 °C
20 % – 80 %, not condensing
free of corrosive fumes, no intensive dust exposure
4.3 Hardware Installation
4.3.1 Room planning
The installation must be performed on a sustainable and vibration-free support. A minimum
distance of 20 cm to the wall is required for a free cooling air circulation.