2042480-001 C Dash 2500 Patient Monitor 8-5
Heart Rate/Pulse: HR/Pulse Menu
Adjust QRS volume
This option lets you adjust the volume of the tone that sounds each time a QRS
complex is detected. If you set the volume to zero, no tone will sound.
Adjust limits
The adjustable and auto-set alarm limits are shown in the table. Choose hi to
change the high alarm limit and lo to change the low alarm limit. The system
maintains appropriate separation between the high and low limits. Choose
auto-set to use the patient’s current readings in setting the high and low limits.
Advanced settings
Limit alarms priority
This option lets you choose the priority of the limit alarms.
Change color based on source?
This option lets you change the color of the HR/Pulse vital signs information on
the Monitor screen to match with the HR/Pulse source parameter color. Choose
yes to have the color appear based on the source. Choose no to select individual
HR/Pulse color.
Select HR/Pulse’s color
This option lets you change the color of this parameter’s vital signs that appear
on the Monitor screen.
HR/Pulse Alarm Limits
Heart Rate/ Pulse
Adjustable
Auto-set
choice: 20%, 30%, 40%
Low High Low High
adult/ped/neonate 30-245 (bpm) 35-250 (bpm) current reading
minus chosen
percentage
current reading
plus chosen
percentage

Table of Contents

Other manuals for GE Dash 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 2500 and is the answer not in the manual?

GE Dash 2500 Specifications

General IconGeneral
BrandGE
ModelDash 2500
CategoryMedical Equipment
LanguageEnglish

Related product manuals