2042480-001 C Dash 2500 Patient Monitor 11-17
NIBP: Specifications
Specifications
GE Healthcare Patents
4,638,810; 5,052,397; 4,754,761; 5,170,795; 6,188,407; 5,357,970; 5,704,362;
5,680,870; 5,518,000; 6,893,403; 6,423,010; 6,358,213; 5,704,362; 5,579,776 and
international equivalents. US patents pending.
Specifications
Method Oscillometric with step deflation
Modes Manual, automatic, stat
BP Measurement Ranges
Systolic 30 to 290 mmHg (adult/pediatric)
4.0 to 38.7 kPa (adult/pediatric)
30 to 140 mmHg (neonate)
4.0 to 18.7 kPa (neonate)
MAP 20 to 260 mmHg (adult/pediatric)
2.7 to 34.7 kPa (adult/pediatric)
20 to 125 mmHg (neonate)
2.7 to 16.7 kPa (neonate)
Diastolic 10 to 220 mmHg (adult/pediatric)
1.3 to 29.3 kPa (adult/pediatric)
10 to 110 mmHg (neonate)
1.3 to 14.7 kPa (neonate)
Resolution 1 mmHg
Accuracy
Meets AAMI/ANSI standard SP10:1992
Initial Cuff Inflation Pressure 135 ± 15 mmHg default; user selectable (adult/pediatric)
100 ± 15 mmHg default; user selectable (neonate)
Maximum Determination Time 120 s (adult/pediatric)
85 s (neonate)
Over Pressure Monitor 300 to 330 mmHg (adult/pediatric)
150 to 165 mmHg (neonate)
Pulse Rate When NIBP is the source, HR values are derived from the pulserate
that is determined by the oscillometric technique of measuring
blood pressure. The rate source field is labeled NIBP.
Adult/Pediatric Range: 30 to 200 bpm (± 3.5% or 3 bpm)
Neonate Range: 30 to 220 bpm (± 3.5% or 3 bpm)