2042480-001 C Dash 2500 Patient Monitor 11-21
NIBP: Alarms
Alarms
NIBP Alarms
Name of Alarms Probable Cause Response
Systolic High
Systolic Low
value is greater than “hi” alarm
limit
value is greater than “lo” alarm
limit
message appears in alarm message field;
parameter label, violated limit (if limits set to
display), and value in violation flash in vital
signs field; NIBP OUT of LIMITS message
appears in SelectBox
Mean High
Mean Low
value is greater than “hi” alarm
limit
value is greater than “lo” alarm
limit
message appears in alarm message field;
parameter label, violated limit (if limits set to
display), and value in violation flash in vital
signs field; NIBP OUT of LIMITS message
appears in SelectBox
Diastolic High
Diastolic Low
value is greater than “hi” alarm
limit
value is greater than “lo” alarm
limit
message appears in alarm message field;
parameter label, violated limit (if limits set to
display), and value in violation flash in vital
signs field; NIBP OUT of LIMITS message
appears in SelectBox
NIBP No
Determination*
insufficient signal message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes
NIBP Level Timeout* remained at one cuff pressure
level for more than 1 minute
due to motion
message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes
NIBP Total Timeout* determination time exceeded 2
minutes due to motion artifact
message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes
NIBP Pump
Timeout*
time for inflation was exceeded
or air leak detected
message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes
NIBP Overpressure* appropriate level or pressure in
the cuff has been exceeded
message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes
* Alarm message appears in SelectBox.