8-6 Dash 2500 Patient Monitor 2042480-001 C
Heart Rate/Pulse: Specifications
Specifications
Specifications
ECG
Heart rate accuracy 30 to 300 bpm
± 3 bpm or 3% of reading, whichever is greater
Time to alarm high heart rate < 10 s per AAMI EC13 - 2002
low heart rate < 10 s per AAMI EC13 - 2002
cardiac standstill < 10 s per AAMI EC13 - 2002
tachycardia waveforms < 10 s per AAMI EC13 - 2002
NELLCOR SpO
2
Range 20 - 250 bpm
Accuracy and Tolerance 20 to 250 bpm ±3 digits
Low Perfusion 20 to 250 bpm ±3 digits
MASIMO SpO
2
Range 25 - 240 bpm
Accuracy and Motion Tolerance
Without Motion 25 to 240 bpm ±3 digits
With Motion normal physiologic range
25 to 240 bpm ±5 digits
Noninvasive Blood Pressure
Range
Adult/ped 30 - 200 bpm
Neonate 30 - 220 bpm
Accuracy ± 3.5%
Alarm Limits
Range 30 - 250 bpm