2042480-001 C Dash 2500 Patient Monitor 4-5
Alarms: Alarm Priorities
Alarm Priority Audible Indications Record on Alarm?
Visual
Indication
Color
Visual Indication
Flashing Rate
Crisis yes, if option chosen red 0.5 Hz
Warning yes, if option chosen red 0.5 Hz
Procedural no red 0.5 Hz
Message none no yellow constant on
OFF none none none none
Default Alarm Priorities
Parameters Crisis Warning Procedural Message
Heart Rate/
Pulse
Limit alarms
priority
ECG VTACH Lead fail,
Replace electrodes
Artifact
NIBP Limit alarms
priority
No determination,
Over-pressure,
Pump timeout,
Total timeout,
Level timeout.
SpO
2
Limit alarms
priority
Lost pulse,
Sensor
disconnected,
Sensor faulty,
Sensor off
RESP Resp
approaching
Limit alarms
priority
Lead fail,
Saturation,
Artifact
TEMP Bad Probe,
Too Hot,
Disconnected