Page #73 background imageLoading...
Page #73 background image
2042480-001 C Dash 2500 Patient Monitor 4-13
Alarms: System Fault Alarm
Default Alarm Limits
Parameter Adult/Ped Neonate
Low High Low High
HR/Pulse 50 bpm (adult)
60 bpm (ped)
150 bpm (adult)
160 bpm (ped)
90 bpm 200 bpm
RESP 6 br/min (adult)
10 br/min (ped)
30 br/min (adult)
60 br/min (ped)
15 br/min 100 br/min
NIBP Systolic: 80 mmHg (adult)
Systolic: 70 mmHg (ped)
Systolic: 200 mmHg (adult)
Systolic: 150 mmHg (ped)
Systolic:
40 mmHg
Systolic:
100 mmHg
Diastolic: 30 mmHg (adult)
Diastolic: 30 mmHg (ped)
Diastolic: 120 mmHg (adult)
Diastolic: 90 mmHg (ped)
Diastolic:
20 mmHg
Diastolic:
60 mmHg
Mean: 40 mmHg (adult)
Mean: 40 mmHg (ped)
Mean: 140 mmHg (adult)
Mean: 100 mmHg (ped)
Mean:
30 mmHg
Mean:
70 mmHg
SpO
2
% Saturation: 90 (adult)
% Saturation: 92 (ped)
% Saturation: 100 (adult)
% Saturation: 100 (ped)
% Saturation: 88 % Saturation: 100