11-10 Dash 2500 Patient Monitor 2042480-001 C
NIBP: NIBP Menu
Adjust limits
The adjustable and auto-set alarm limits are shown in the table. Choose
hi
to
change the high alarm limit and
lo
to change the low alarm limit. You can set high
and low limits individually for systolic, diastolic, and mean values. The system
maintains appropriate separation between the high and low limits. Choose
auto-
set
to use the patient’s current readings in setting the high and low limits.
NIBP Alarm Limits (mmHg)
Patient
Adjustable
Auto-set
choices: 20%, 30%, or 40%
Low High mmHg Low High
Systolic
adult/
pediatric
30 - 285 35 - 290 > 25
<
25
current reading minus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
neonatal 30 - 135 35 - 140 > 25
<
25
current reading minus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
Mean
adult/
pediatric
20 - 255 25 - 260 > 25
<
25
current reading minus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
neonatal 20 - 120 25 - 125 > 25
<
25
current reading minus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
Diastolic
adult/
pediatric
10 - 215 15 - 220 > 25
<
25
current reading minus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
neonatal 10 - 105 15 - 110 > 25
<
25
current reading minus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5