2042480-001 C Dash 2500 Patient Monitor 15-7
Battery: Specifications
previously stated, nickel-metal hydride batteries should not be disassembled,
opened, or shredded. Do not short circuit the terminals.
The nickel-metal hydride battery supplied is recognized by the regulatory
agencies as a "dry battery." As such, it is not subject to regulation and can be
shipped in normal packaging and transported on any mode of transportation
without special handling.
The 3.6 volt nickel-metal hydride, rechargeable, memory backup battery
contains no lead, lithium, or cadmium compounds. Discharge this battery prior
to disposal. Place the battery in packaging which electrically isolates its
contents. Do not puncture or place the battery in a trash compactor. Do not
incinerate the battery or expose it to fire or high temperatures.
Specifications
Specifications
Capacity 8.4 V; 7.0 amp-hr
Battery Life greater than 180 minutes using fully charged internal battery
(NIBP: 5-min auto cycle with adult cuff. ECG, RESP, SpO
2
: Active. TEMP:
predictive mode. Printer: printing 2 waveforms for 1 min every 20 min
at 25 mm/sec.)
Charge time 4 hours maximum with the Monitor switched off
8 hours maximum with the Monitor switched on