13-14 Dash 2500 Patient Monitor 2042480-001 C
MASIMO SET® SpO2: Specifications
MASIMO
®
Patents
5,482,036; 5,490,505; 5,632,272; 5,685,299; 5,758,644; 5,769,785; 6,002,952;
6,036,642; 6,067,462; 6,157,850; 6,206,830; 6,263,222 and international
equivalents.
Specifications: MASIMO
®
Sensor Accuracy
Sensor Model
SpO
2
Range
70% - 100%
LNOP
LNOP-ADT ± 2 digits
LNOP-ADT Long ± 2 digits
LNOP-PDT ± 2 digits
LNOP-NEO ± 3 digits
LNOP-NEO PT ± 3 digits
LNOP-DCI (reusable) ± 2 digits
LNOP-DCSC (reusable) ± 2 digits
LNOP-DCIP (reusable) ± 2 digits
NRI25 (reusable) ± 2 digits
Resolution
Saturation (% SpO
2
) 1%
Pulse Rate (bpm) 1
Low Perfusion Performance
0.02% Pulse Amplitude Saturation (% SpO
2
) ±2 digits
and% Transmission >5% Pulse Rate ±3 digits
Interfering Substances Carboxyhemoglobin may erroneously increase
readings. The level of increase is approximately
equal to the amount of carboxyhemoglobin
present. Dyes, or any substance containing
dyes, that change usual arterial pigmentation
may cause erroneous readings.
Sensor Light Source
Wavelength Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation Infrared: 22.5 mW (max)
Red: 27.5 mW (max)