2042480-001 C Dash 2500 Patient Monitor 4-11
Alarms: View Event History Menu
crisis alarm generated by the ECG parameter
crisis alarm generated by the HR/Pulse parameter
Freeze hardkey is pushed
Events with stored waveforms are indicated by an asterisk (*) in the Wave
column. A maximum of 12 patient events may be stored with a frozen
waveform. If 12 events exist, the next event with a frozen waveform will replace
the oldest stored waveform and event. If this older event is still one of the most
recent 200 patient events, it continues to appear in the Patient Event History
window without a stored waveform or an asterisk (*).
Choose event history
This option lets you choose to view either the Patient Event History or System
Event History window.
view older events
This option lets you view previous events. Each time you push the Trim Knob, a
new screen of events is displayed.
view newer events
This option lets you access the most recent events, including any events stored
while the Event History window is open. Each time you push the Trim Knob, a
new screen of events is displayed.
print this page
This option lets you print the list of events currently on the screen. It is available
when the printer is not busy and the crisis events and frozen waveforms option
is closed.
print all pages
This option lets you print all events being displayed in the Event History window
(either Patient Events or System Events). It is available when the printer is not
busy and the crisis events and frozen events option is closed.
NOTE: If any new events occur while a page of the oldest events is being
viewed/printed, then an equal number of the oldest events will be
removed from memory and no longer available for printing.
crisis events and frozen waveforms
This extended menu is available only when the Patient Event History window is
chosen and there is at least one stored event and the freeze feature is not active.

Table of Contents

Other manuals for GE Dash 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 2500 and is the answer not in the manual?

GE Dash 2500 Specifications

General IconGeneral
BrandGE
ModelDash 2500
CategoryMedical Equipment
LanguageEnglish

Related product manuals