14-10 Dash 2500 Patient Monitor 2042480-001 C
Alaris® Turbo Temp®: Probe Cleaning and Disinfection
Probe Cleaning and Disinfection
Do not immerse predictive temperature probes. The probe can be cleaned with a
solution of 10% bleach in water. Use a cloth or sponge-just damp, not wet-and
avoid getting any liquid into the interior of the probe.
Calibration
The temperature parameter does not require calibration.
Specifications
Alaris
®
Patents
U.S. D300,728, D300,909. Other pending patents.
Specifications
Scale ° Fahrenheit (F)
° Celsius (C)
Predictive Mode
Range 96.0°F (35.6°C) to 106.0°F (41.1°C)
Resolution 0.1°F (0.1°C)
Monitor Mode
Range 80.0°F (26.7°C) to 108.0° F (42.2°C)
Accuracy ± 0.2°F (± 0.1°C) (when tested in a calibrated liquid bath; meets
ASTM E1112, Table 1, in range specified)
Resolution 0.1°F (0.1°C)
Probes CAUTION:
Use only Alaris
®
probes and probe covers. The size, shape, and
thermal characteristics of the probe covers can affect the
performance of the instrument. Inaccurate readings or retention
problems may occur unless Alaris
®
probes and probe covers are
used.
Determination Time Approx. 10 seconds, typical