12-12 Dash 2500 Patient Monitor 2042480-001 C
NELLCOR® OXIMAX® SpO2: Specifications
NELLCOR
®
Puritan Bennett, Inc. Patents
4,621,643; 4,653,498; Re.35,122; 4,700,708; 4,770,179; 4,802,486; 4,869,254;
4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,421,329; 5,485,847; 5,533,507;
5,577,500; 5,803,910; 5,853,364; 5,865,736; 6,083,172 and international
equivalents.
OxiCliq
®
OxiCliq A ±2.5 digits
OxiCliq P ±2.5 digits
OxiCliq N
(Adult)
±2.5 digits
OxiCliq N
†
(Neonate)
±3.5 digits
OxiCliq I ±2.5 digits
Reusable Sensor Models
D-YS (Infant to Adult) ±3 digits
D-YS (Neonate) ±4 digits
D-YS & D-YSE ±3.5 digits
D-YS & D-YSPD ±3.5 digits
DS-100A ±3 digits
OXI-A/N (Adult) ±3 digits
OXI-A/N (Neonate) ±4 digits
OXI-P/I ±3 digits
Neonatal Sensor Accuracy When sensors are used on neonatal subjects as
recommended, the specified accuracy range is increased
by ±1 digit, as compared to adult usage, to account for
the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood. For example, MAX-N
accuracy on neonates is ±3 digits, rather than ±2 digits.
Sensor Light Source
Wavelength Infrared: 890 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation Infrared: 22.5 mW (max)
Red: 30 mW (max)
†
The MAX-N, D-YS, OXI-A/N, and OxiCliq N were tested on patients >40 kg.
‡
The accuracy specification has been determined between saturations of 80%-100%.
Specifications: NELLCOR
®
OXIMAX
®
Sensor Accuracy