Software Installation &Maintenance 6-11
3. Click [OK]. A progress bar appears and the setup data of the selected item is exported to the
specified path.
6.6.1.2 Restore the Setup Data
1. Click [Import] to open the Load Data dialogue box.
2. Select the path to import the data. Select the path to import all data or some of the data;
3. Click [OK]; a progress bar appears and the setup data is imported to the specified module.
NOTE: If selecting [Load Factory], the settings are restored to the factory defaults, except for region
preset, admin, DICOM setting and network setting (including iStorage).
6.6.2 Patient Data Backup and Restoration
6.6.2.1 Patient Data Backup
1. Press [iStation] on the keyboard to open the iStation dialogue box;
2. Click [Select All Exams] to select all data or select the desired data one by one;
3. Select the information. Click [Backup Exam] to open Backup Patient Record dialogue box.
Select the desired storage device (recorder or USB disk). Click [Backup]. Back up the selected
data automatically.
4. Select whether to encrypt backup exams for USB only: input the password and confirm
password in the field box, and then click [Backup]. A "Patient.7z" compressed package is
backed up to the USB device, and you need to input the password to open the package.
Note: If the password is forgotten, you cannot open the backup package.
6.6.2.2 Restore Backup Patient Data
1. Press [iStation] on the keyboard to open the iStation dialogue box;
2. Select the drive which contains the patient data. Click [Select All Exams] to select all data or
select the patient data to be restored one by one. Click [Restore] to restore the patient data
from the current drive to the patient database.
6.7 Probe Check
This function enables users to check if a transducer element is in malfunction, so as to evaluate the
transducer performance.
Log on to the system with the Service account, Press [F10 Setup] to select [Maintenance]
→[Setup]→[Probe Check] to enter Probe Check screen. If a transducer element is in
malfunction, it is displayed as a red spot. If a transducer element functions normally, it is
displayed as a green spot.
You can click [Export] to export the result to an external device.
Note:
1. Probe check is provided for reference only, not for confirming a diagnosis.
2. Each time a transducer is connected to the system or activated, the users are
recommended to implement probe check of transducer performance.