3-32 System Installation
5. Click wireless network logo on the lower right corner of the Windows screen. Select the
desired EAP network, and click [Connect].
If the EAP network cannot be connected, operate according to the prompts on the
screen
6. Double-click the tool “BackToDoppler” on the desktop to return to the Doppler interface.
7. Click [Disable ms-setting] on the maintenance screen. Save the settings. The system is
automatically powered off.
8. Power on the system. Select the desired EAP network on the wireless network
connection menu, and click [Connect].
To remove the added EAP network, perform steps 1-3. Open Startup menu on the
Windows interface. Select [Settings]→[Network & Internet]→[Wi-Fi] →[Manage known
networks], select the desired EAP network, and click [Forget].
After clicking [Enable ms-setting], you can make settings for the operating system on
Windows interface. After the settings, click [Disable ms-setting] on the maintenance
screen to disable the function immediately. Otherwise, the performance or operation
of the ultrasound system may be affected.
3.7.7 DICOM/HL7 Preset
Only if DICOM basic option is configured, [DICOM Preset] is available.
1. Click [DICOM/HL7] to open the DICOM Preset screen. Enter the AE Title of the ultrasound
system, port and PDU according to the actual situation.
2. DICOM Server Setting
1) Enter the device name and the IP address.
2) You can ping other machines to verify connection after entering the correct IP address by
clicking [Ping].
3) Click [Add] to add the server to the list if the connection works normally.
The following is an example: