12-10 Care and Maintenance
12.3.4 Mechanical Safety Inspection
Mechanical safety inspection is mainly used to check mechanical strength and mechanical function
of the key assembly of ultrasonic system. The mode of test evaluation mainly is: Perform the
evaluation by means of visual check and operating check, if the check result cannot pass, the
system is in abnormal status now. Stop using the system and adopt proper measures. The test flow
is as following:
Check the handle
Check control panel And
supports
Check the turning and
moving of monitor
Check other mechanics
Check the connection of the
casters
Check the casters
Check the turning and lifting
Fig 12-4 Mechanical Safety Inspection Flow