9-94 Structure and Assembly/Disassembly
9.3.39 Magnetic Generator Trolley
The disassembly tool: cross-headed screwdriver (M3, M4), inner hex spanner (M5).
9.3.39.1 Sliding Pole Assembly
1. Hold the sliding pole assembly on one hand, and loosen the sliding pole anticlockwise on another
hand. Lift the sliding pole straight up.
Note:
In case of the scratch to the antiskid thermal shrinkage pole, loosen the adjusting
knob completely.
9.3.39.2 Connecting Base Assembly
1. Shake the storage tray gently with two hands holding the edge of the storage tray. Lift the
storage tray up.
Adjusting knob
S
liding pole assembly
()