EasyManuals Logo
Home>Mindray>Diagnostic Equipment>Eagus R9

Mindray Eagus R9 User Manual

Mindray Eagus R9
365 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #319 background imageLoading...
Page #319 background image
Troubleshooting of Regular Malfunctions 13-11
13.7.3 Troubleshooting of the Monitor
No. Fault Description Cause Analysis Measure
1
Control panel powered on
normally;
No display (blank screen) on
the LCD;
Green indicator of the
display is on.
Monitor error
Replace monitor
package
2 Control panel powered on
normally; one of indicators
turns to blue.
No display (blank screen) on
the LCD;
Display indicator is off.
Abnormal power supply of the
monitor and the monitor error
Check the power supply
for the monitor or
replace the monitor
material package
3 Control panel powered on
normally;
“No Signal” displays on LCD;
Orange indicator of the
display is on.
Normal touch screen, the image
appears on the external monitor.
Monitor signal transmitting plate
or PC carrier board errors.
Replace the plate or PC
carrier board.
Normal touch screen, the monitor
image appears. PC carrier board
error
Replace PC carrier
board
Abnormal touch screen. CPU
error.
Replace CPU module
4 LCD works normally;
Touch screen blank.
Abnormal power supply of the
touch screen and the touch
screen error
Replace the power
supply of the touch
screen or touch screen
assembly
5
LCD works normally;
“No Signal”
CPU module error or PC carrier
board error
Replace CPU module or
PC carrier board
13.8 ECG Module Failure
13.8.1 Related Modules or Boards
No. Description Remarks
1 ECG assembly /
2
Engine board
/

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Eagus R9 and is the answer not in the manual?

Mindray Eagus R9 Specifications

General IconGeneral
BrandMindray
ModelEagus R9
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals