Chapter 9: Installing, Removing, and Replacing Components
Removing and Replacing the System Control Board
Quantum Scalar i40 and Scalar i80 User’s Guide 285
3 Complete the Setup Wizard on the operator panel. If you make a
mistake or do not complete all the screens, you can change the
configuration later using the operator panel or Web client Setup
menus.
Note: You cannot update the host name or time zone on the
operator panel. If the displayed host name or time zone is
incorrect, change it later using the Web client (see
Step 8
on page 285).
4 Take note of the library’s IP address (on the operator panel, select
Reports > Network Settings).
5 Access the Web client using the IP address and log in using the
default user name and password.
•
6 Optionally, install tape drive firmware for autoleveling. The library
firmware contains baseline tape drive firmware for autoleveling.
Verify your tape drive firmware is at the level you expect. If not:
a Upload the tape drive firmware you want to use. From the Web
client, select Tools > Drive Operations > Upload/remove tape
drive firmware for autoleveling.
b Unload all tape cartridges from the tape drives on which you
want to autolevel firmware.
c Initiate autoleveling by doing one of the following:
• Restart the library (from the operator panel, select Actions
> Shutdown > Restart Library.
• Restart all the tape drives you want to autolevel, as follows:
From the Web client, select Tools > Diagnostics. Confirm
you want to take the partitions offline. Select Drives >
Drive Tests > Drive Reset. Select the applicable partitions
and tape drives, then click Apply, then click OK.
7 If you changed the IP address from what it was originally, update
host and other applications that access the library.
8 Update the library name, time zone, and any other configuration
settings, if required, using the Web client Setup menus.