EasyManuals Logo
Home>ABB>Laboratory Equipment>PGC2000

ABB PGC2000 User Manual

ABB PGC2000
400 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #116 background imageLoading...
Page #116 background image
4-35 2000-OSM, F1
will reduce the classification to Nonhazardous/General Purpose. Since the Y and Z Purge systems
are identical except for nomenclature, the following explanation applies to both.
The purge air enters the analyzer system through a pressure regulator, an orifice, and a gauge used
to set purge flow, and then into the electronics housing where purge air flow is maintained until a fault
condition exists.
When a purge air alarm condition exists, the GCC will set an *** ALARM *** signal display on the
screen and cause a purge alarm contact to switch. A Y or Z Purge alarm alerts you to the condition,
but it does not turn off the analyzer.
NOTE
It is the customer’s responsibility to connect the purge alarm contact to a visual or
audible annunciator located in a constantly monitored area, as referenced in NFPA
496 (2003) and IEC 60079-2:2001.
Purge time depends on the application. Refer to the analyzer label or Data Package for specific
analyzer purge time.
X Purge Operation
While X Purge (see Figure 4-36 for features and Figure 1-1 for location) is an optional feature, some
installations specifying certification require it. X Purge reduces risk by two levels and turns off the
analyzer when an alarm occurs.
CUSTOMER
POWER
CONNECTIONS
L1 (HIGH)
L2 (NEUT)
GND
OVERRIDE PUSH BUTTON
1 2 3 4
TB70
Figure 4-36. X PURGE WITH COVER REMOVED
When you turn the power ON, X Purge begins monitoring the electronics housing (GCC and Purge
Air Panel) air purge pressure. When the air purge pressure reaches the specified level, the X Purge
starts a timed cycle. If the air purge pressure remains at the specified level during the timed cycle, X
Purge will supply power to the equipment when it completes the timed cycle. If the air purge pressure
drops before the X Purge completes the timed cycle, X Purge resets the timer and starts over. It
continues to reset the timer until the analyzer achieves and maintains the correct pressure for a
complete timed cycle. The X Purge will not supply power to the analyzer until it successfully
completes the timed cycle.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB PGC2000 and is the answer not in the manual?

ABB PGC2000 Specifications

General IconGeneral
BrandABB
ModelPGC2000
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals