EasyManuals Logo
Home>ABB>Laboratory Equipment>PGC2000

ABB PGC2000 User Manual

ABB PGC2000
400 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
2000-OSM, F1 2-13
Checking Backflush Flow
1. Connect the flow-measuring device to the Backflush Vent and verify that Valves One and Two are
OFF (see Figure 2-8).
2. Adjust the Backflush Vent restrictor to the flow rate stated in the Data Package.
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
9
10
10
CARRIER
IN
COLUMN
PRESSURE
ADJ
PI
AIR IN
AIR OUT
VALVE “OFF”
AIR IN
AIR OUT
VALVE “OFF”
SAMPLE
VENT
SAMPLE
IN
ANALYTICAL
FLOW ADJ
BACKFLUSH
VENT
COLUMN 1
SAMPLE
LOOP
COLUMN
2
COLUMN
3
DETECTOR
SELECTOR
FLOW ADJ
SELECTOR
VENT
CELL
VENT
Figure 2-8. CHECKING BACKFLUSH FLOW

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB PGC2000 and is the answer not in the manual?

ABB PGC2000 Specifications

General IconGeneral
BrandABB
ModelPGC2000
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals