2-34 2000-OSM, F1
13. When you have completed and verified your entries, press the F2 (Exit and Update) soft key to
save your entries.
******* VistaNET Essential Addresses *******
Name:
[GCC10 ]
Tag: GCC10
Board Address: 0.10
Domain 1: 0.1
Domain 2: 0.2
VNSA Address: 0.1.0.199
Router Address: 0.1.0.199
Exit
Escape and
Update
F1 F2 F3 F4
Figure 2-20. VistaNET ESSENTIAL ADDRESSES SCREEN
Setting Access Levels
Next you must enter the appropriate access level. There are three levels of remote access (i.e., via
VistaNET) granted by the analyzer:
1. Net Locked: no remote modification of the analyzer's tables is permitted. It requires a
SUPERVISOR VNSA access level to modify any of the analyzer's tables from the front panel (see
the applicable VistaNET Installation Guide for more information).
2. Restricted: remote modification of tables is limited to users that have SUPERVISOR or NORMAL
access level.
3. Unrestricted: remote modifications can be performed by any user.
You set the remote access level for an analyzer from the front panel as follows: