1-4 2000-OSM, F1
Power
WARNING
The Neutral and Ground (Earth) connections to the PGC2000 must be at earth (0
Volts) potential. Failure to maintain earth (0 volts) potential at these connection
points constitutes a serious safety hazard.
Voltage and Frequency: 100 VAC (+15, -6 VAC), 50/60 Hz ± 10%
120 VAC ± 10%, 50/60 Hz ± 10%
230 VAC ± 10%, 50/60 Hz ± 10%
Power Consumption: 1,200 VA Startup, 900 VA Steady-State Operation
(Typical, varies with installed options)
Instrument Air
Supply Connection: 3/8 inch tube, minimum
Supply Pressure: 60 psig (414 kPa, 4.14 bar) minimum
100 psig (689 kPa, 6.89 bar) maximum
Quality: Instrument grade: Clean, Oil Free and -34° C, (-30° F) dewpoint
Flow Rates, all purge types:
Start-up Purge: 7.6-8.5 ft
3
/min (214-242 L/min) at 60°F (20°C)
Steady State Purge : 4.5-5.2 ft
3
/min (127-147 L/min) at 60°F (20° C)
Analytical Detectors
Standard Detectors: Thermal Conductivity, Flame Ionization, Flame Photometric,
Inter-column Thermal Conductivity
Configuration: Up to two simultaneous detectors, each with capability for
sequential inter-column TC detector (one FPD per analyzer)
Third-party detectors: Consult Factory for availability
Isothermal Analytical Oven
Oven Liner: Stainless Steel
Internal Dimensions: 15.3 in. W x 20.4 in. H x 9.0 in. D
(390 mm W x 520 mm H x 230 mm D)
Number of Valves: Standard provisions for five gas samples or column switching
valves in oven; standard provisions for two external liquid sample
valves (consult factory for special requirements)
Columns: 1/16, 1/8, 3/16 inch, Packed Stainless or Fused Silica Capillary
Heat: Forced Air