2-18 2000-OSM, F1
b. On the Chroma Board screen, verify that the FPD Temp Zone is the same as that specified in
the Data Package (see Figure 2-10).
c. On the Chroma Board screen, observe the "Ignitor" line. "Manual" should be underlined.
d. Press the F1 (Exit) soft key.
e. On the Manual Control Mode Screen, press the F4 (Press. & Temp. Control) soft key.
f. On the Pres and Temp Control screen, cursor to TEMP CONTROL and press the F2 (Manual
Temp Control) soft key.
***** Chroma Board *****
Detector: FPD
Flame: Lit Ignitor: Off H2: None
Hardware Zero: 5.10mV ( 4.7%)
Software Zero: - 3.18mV
Detector Reading: 1.98mV
Ignitor: Manual Auto Auto Timeout
Timeout: 0 seconds
FPD Temp Zone: 0
Manually Manual
Exit Graphics Zero Ignitor
Baseline On
F1 F2 F3 F4
Figure 2-10. CHROMA BOARD SCREEN