EasyManuals Logo
Home>ABB>Laboratory Equipment>PGC2000

ABB PGC2000 User Manual

ABB PGC2000
400 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
2000-OSM, F1 2-9
5. Refer to Figure 2-4 (when EPC is installed, the Pressure Control Panel will be blank). The purge
air and heater air pressures should be as specified on the label below the corresponding
regulator.
Purge Air Panel Pressure Control Panel
Figure 2-4. ANALYZER ADJUSTMENTS
6. Turn on carrier gas at its source.
7. Refer to the Data Package to ensure that the specified carrier gas is provided at the specified
pressure to each CARRIER inlet.
8. At the external circuit breaker or switch turn the power ON.
9. For EPC-equipped analyzers, refer to the Manual Pressure Control Mode screen. Verify that
carrier and gas pressures are the same as those specified in the Data Package. If necessary,
adjust each set point (SP) until it is the same as the Data Package value.
10. If the analyzer has mechanical regulators, verify from the Pressure Control Panel gauge that
carrier and pressures are the same as that specified in the Data Package. If necessary, adjust
each regulator until its gauge value is the same as the Data Package value.
11. Verify any other gases in the same manner as Carrier #1.
12. If a leak occurs during start-up, perform the leak check described in "Leak Check."
13. Allow sufficient time for oven warm-up. The oven should stabilize at the temperature indicated in
the Data Package. If adjustments are required, adjust the Digital Temperature Controller as
described in "DIGITAL TEMPERATURE CONTROLLER" in Section 4.
14. Proceed to "MATCHING COLUMN FLOWS."
Leak Check
If a leak occurs during initial turn-on, perform a simple leak check on the analyzer section as follows:
1. Plug all carrier gas vents and pressurize to 30 psig. Shut off the carrier supply.
2. Note pressure drop-off on the carrier supply and column pressure gauges. A visual decline is
excessive.
3. Starting at the inlet to the detector and working back to the carrier gas supply, plug the exit of
each valve or column to locate the leak.
4. Use “Snoop®“ or other suitable leak detection solution to check for leaks.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB PGC2000 and is the answer not in the manual?

ABB PGC2000 Specifications

General IconGeneral
BrandABB
ModelPGC2000
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals