10-14 2000-OSM, F1
Comments: time! is the time to convert to a percent off.
time_vs_bp_table% is the Time vs. Boiling Point Table (1-8) to use for the
conversion.
Example: 0010 PERCENT! = TIME_TO_OFF(100.0,1)
Returns the percent off that occurred at 100.0 seconds.
GET PERCENT OFF TABLE Command
Description: Returns the percent off column of one of the eight Time vs. Boiling Point
tables (Sim Dis analyzers only).
Syntax: GET_POFF_TBL(array!,table%)
Comments: array! is a previously dimensioned array that is the destination of the
Percent Off Table.
table% is the table number (1-8).
Example: 0010 N = LEN_POFF_TBL(1)
0020 DIM A!(N)
0030 GET_POFF_TBL(A!,1)
Gets the Percent Off Table size, dimensions an array to hold it, copies the
table into array A!.
GET PRESSURE OF A ZONE Command
Description: Returns the actual pressure of the specified zone.
Syntax: GET_PRESS(zone%)
Comments: zone% is the zone number (1-5).
Example: 0010 Z1! = GET_PRESS(1)
Returns the actual pressure of zone 1.
GET RESPONSE FACTOR Function
Description: Returns the component’s response factor.
Syntax: GET_RF(method%,component%)
Comments: method% is the method table (1-8).
component% is the component number for which the response factor is to
be retrieved.
Example: 0010 A! = GET_RF(1,2)
Gets the response factor of component two from analysis method one and
assigns it to floating point variable A!.
GET STREAM ASSIGNMENT Function
Description: Returns the analysis method assigned to a stream.
Syntax: GET_ASSIGN(stream%)
Comments: stream% is the stream number (from 1 to the maximum number of
streams).
Example: 0010 M = GET_ASSIGN(5)
The variable M gets the table number of the method assigned for the
analysis of stream #5.
GET TEMPERATURE FOR A PERCENT OFF Command
Description: Returns the temperature for a given percent off using a specified Time vs.
Boiling Point Table (Sim Dis analyzers only).
Syntax: GET_TEMP(percent_off!,time_vs_bp_table%)