Do not touch the patient and the Signal Input/Output lines (for example, USB port) at the same
time.
2.5.2 Environmental Conditions for Operation
Temperature: 18˚C to 30˚C resp. 64˚F to 86˚F
Humidity: 30% to 80% RH, no condensation
Barometric pressure: 700 to 1060 hPa
Light conditions: natural & artificial light source*
Maximum operating altitude: 3000m; depending on the properties of the connected electronic
devices the maximum operating altitude is limited to the altitude stated
in the corresponding user manual of the connected electronic device
Pollution degree: 2
Overvoltage category: II
Material group: IIIB
Total audible noise emission: <55dB
Note
* Bright light could impact readability of screen.
Ultrasound systems are highly sensitive medical instruments that can easily be damaged by
improper handling. Use care when handling and protect from damage also when not in use.
DO NOT use a damaged or defective ultrasound system. Failure to follow these precautions
can result in serious injury and equipment damage.
This equipment is not to be used during transportation (e.g. ambulance cars, aircrafts).
Thermal Safety. Maintaining a safe thermal environment for the patient has been a design
priority at GE Healthcare. Software settings limit the power dissipated for the ultrasound
transducer as well as the motor-drive to values low enough to ensure that operating
temperatures stay below 43˚C.
Not to be used in sterile environment.
This equipment must not be used in oxygen enriched atmosphere or in the presence of
inflammable gases (e.g. anesthetic gases).
The use of the system outside the described conditions or intended use, and disregarding
safety related information is considered as abnormal use. The manufacturer is not liable for
damage caused by abnormal use of the device!
Use for diagnostic purposes only.
Safety
2-8
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4

Table of Contents

Other manuals for GE Voluson S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson S6 and is the answer not in the manual?

GE Voluson S6 Specifications

General IconGeneral
BrandGE
ModelVoluson S6
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals