11.2.2.11 HR (Heart Rate)
To measure the Heart Rate (HR) in M Mode:
1. Press the [Calc] key on the control panel.
2. Select the desired item. For example: select [Left Renal Artery].
3. Select the [HR] key on the control panel. A line appears on the screen.
4. Move the line to the start point of the period and press the right or left trackball key [Set].
The second line appears.
5. Select the number of heart rate cycles necessary for measurement with the rotary control
below the control panel.
6. Move the second line to the end point of the period (according to the selected number of
heart rate cycles to be measured).
7. Press the right or the left trackball [Set] key again. The Heart Rate is displayed.
11.2.2.12 Abdomen Calculations in Spectral-Doppler Mode
There are many possibilities to measure different vessels in Spectral-Doppler mode:
Auto Trace
(chapter
'Auto Trace'
on page 11-31
)
Measurements and Patient Worksheets (Reports)
11-30
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4

Table of Contents

Other manuals for GE Voluson S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson S6 and is the answer not in the manual?

GE Voluson S6 Specifications

General IconGeneral
BrandGE
ModelVoluson S6
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals