For example: VOLUSON
®
TRANSVAGINAL TRANSDUCER RIC5-9W-RS
1. Medical probe-sheath
4.6.1 Probe Usage
For details about connecting, activating, deactivating, disconnecting, transporting and storing
the probes, refer to: 'Transducer Connection'
on page 4-5
and Probe/Program Selection
'Probe/Program Selection'
on page 4-7
.
4.6.1.1 Coupling Gels
Do not use unrecommended gels (lubricants). They may damage the probe and void the
warranty.
Applying:
In order to assure optimal transmission of energy between the patient and probe, a conductive
gel or couplant must be applied liberally to the patient where scanning will be performed.
Precautions:
Coupling gels should not contain the following ingredients, as they are known to cause probe
damage:
Methanol, ethanol, isopropanol, or any other alcohol-based product
Mineral oil
Iodine
Lotions
Lanolin
Aloe Vera
Olive Oil
Methyl or Ethyl Parabens (para hydroxybenzoic acid)
Dimethylsilicone
Operating the System
4-6
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4

Table of Contents

Other manuals for GE Voluson S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson S6 and is the answer not in the manual?

GE Voluson S6 Specifications

General IconGeneral
BrandGE
ModelVoluson S6
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals