4.8.3.2 Patient Information – Obstetrics (OB)
Application Data:
LMP Enter the date of the last menstrual period using the selected format
(e.g., mm-dd-yyyy).
Note: The first day of the last period must be entered.
DOC Enter the date of conception.
EDD Enter the estimated date of delivery, GA is calculated automatically.
Can be changed to 41 weeks.
GA GA will be calculated automatically after entering the date of the last
menstrual period (LMP) or the estimated date of delivery (EDD).
“LMP” in brackets denotes that GA was calculated. For future exams,
the GA will be calculated from the LMP.
When entering the Gestational Age (GA), EDD and DOC are
calculated automatically.
“CLIN” in brackets denotes that the GA was tested. For future exams,
the GA will be calculated from the GA.
Gravida Enter the patient’s history of pregnancies.
Para Enter the patient’s history of live births.
Aborta Enter the patient’s history of abortions.
Ectopic Enter the patient’s history of ectopic pregnancies.
Fetus # Enter the fetus number (e.g., at twins “2”).
When entering LMP, the GA and EDD fields automatically show the calculation results.
When entering GA only EDD is calculated; when entering EDD only GA is calculated.
Operating the System
4-16
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4

Table of Contents

Other manuals for GE Voluson S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson S6 and is the answer not in the manual?

GE Voluson S6 Specifications

General IconGeneral
BrandGE
ModelVoluson S6
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals