16.12.6 Internal Battery
Type CR2450
Note Engraved “+” marking shall be on top position.
16.13 Guidance and manufacturer´s declaration
Guidance and manufacturer´s declaration - electromagnetic emissions
The Voluson® S6/S8 is inteded for use in electromagnetic enviroment specified below. The customer or
the user of the Voluson® S6/S8 should assure that it is used in such an enviroment.
Emission test Compliance Electromagnetic enviroment - guidance
RF emissions
CISPR 11
Group 1 The Voluson® S6/S8 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The Voluson® S6/S8 is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes,
provided the following warning in heeded:
WARNING: The Voluson® S6/S8 is intended for
use by healthcare professionals only. The
Voluson® S6/S8 may cause radio interference or
may disrupt the operation of nearby equipment.
It may be necessary to take mitigation measures,
such as re-orienting or relocating the system or
shielding the location.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer´s declaration - electromagnetic emissions
The Voluson® S6/S8 is inteded for use in electromagnetic environment specified below. The customer or
the user of the Voluson® S6/S8 should assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance level Electromagnetic environment- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 2, 4, 6kV contact ± 2, 4, 6kV contact Floors should be wood,concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
± 2, 4, 8kV air ± 2, 4, 8kV air
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power
supply lines
± 2kV for power
supply lines
Mains power quality should be that of a
typical commercial or hospital
environment.
±1kV for input/
output lines
±1kV for input/
output lines
Surge
IEC 61000-4-5
± 1kV differential
mode
± 1kV differential
mode
Mains power quality should be that of a
typical commercial or hospital
environment.
± 2kV common
mode
± 2kV common
mode
Technical Data / Information
Voluson® S6/S8 Basic User Manual