EasyManuals Logo
Home>Keithley>Measuring Instruments>2600B

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #109 background imageLoading...
Page #109 background image
Series 2600B
System SourceMeter® Instrument Reference Manual Section 2:
General operation
2600BS-901-01 Rev. C / August 2016 2-85
Range programming
Range commands
The following tables summarize commands necessary to control measure and source ranges. See
the TSP command reference (on page 7-1
) for more details about these commands.
Measure range commands*
Command** Description
smuX.measure.autorangei = smuX.AUTORANGE_ON
Enable current measure autorange.
smuX.measure.autorangei = smuX.AUTORANGE_OFF
Disable current measure autorange.
smuX.measure.autorangev = smuX.AUTORANGE_ON
Enable voltage measure autorange.
smuX.measure.autorangev = smuX.AUTORANGE_OFF
Disable voltage measure autorange.
smuX.measure.lowrangei = lowrange
Set lowest I measure range for
autorange.
smuX.measure.lowrangev = lowrange
Set lowest V measure range for
autorange.
smuX.measure.rangei = rangeval
Select manual current measure range.
smuX.measure.rangev = rangeval
Select manual voltage measure range.
* See Available ranges (on page 2-81)
** smuX: For Models 2601B, 2611B, and 2635B, this value is smua (SMU Channel A); for Models 2602B, 2604B, 2612B,
2614B, 2634B, and 2636B, this value can be
smua
(for SMU Channel A) or
smub
(for SMU Channel B).
Source range and limit commands*
Command** Description
smuX.source.autorangei = smuX.AUTORANGE_ON
Enable current source autorange.
smuX.source.autorangei = smuX.AUTORANGE_OFF
Disable current source autorange.
smuX.source.autorangev = smuX.AUTORANGE_ON
Enable voltage source autorange.
smuX.source.autorangev = smuX.AUTORANGE_OFF
Disable voltage source autorange.
smuX.source.limiti = level
Set voltage source current limit
(compliance).
smuX.source.limitv = level
Set current source voltage limit
(compliance).
smuX.source.limitp = level
Set source power limit (compliance).
smuX.source.lowrangei = lowrange
Set lowest I source range for autorange.
smuX.source.lowrangev = lowrange
Set lowest V source range for autorange.
smuX.source.rangei = rangeval
Select manual current source range.
smuX.source.rangev = rangeval
Select manual voltage source range.
* See Available ranges (on page 2-81)
** smuX: For Models 2601B, 2611B, and 2635B, this value is smua (SMU Channel A); for Models 2602B, 2604B, 2612B,
2614B, 2634B, and 2636B, this value can be
smua
(for SMU Channel A) or
smub
(for SMU Channel B).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600B and is the answer not in the manual?

Keithley 2600B Specifications

General IconGeneral
BrandKeithley
Model2600B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals