7: TSP command reference Series 2600B System SourceMeter® Instrument
7-332 2600BS-901-01 Rev. C / August 2016
Bit B7 B6 B5 B4 B3 B2 B1 B0
(2
) (2
) (2
) (2
) (2
) (2
) (2
) (2
)
Binary value 0/1 0/1 0/1 0/1 0/1 0/1 0/1 0/1
Example
status.questionable.enable =
status.questionable.OTEMP
Uses a constant to set the OTEMP bit of
the questionable status enable register.
Also see
Questionable Status Registers (on page E-13)
status.questionable.calibration.*
This attribute contains the questionable status calibration summary register set.
Type TSP-Link accessible Affected by Where saved Default value
.enable (RW) Yes Status reset Not saved 0
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B: 6
Usage
questionableRegister = status.questionable.calibration.condition
questionableRegister = status.questionable.calibration.enable
questionableRegister = status.questionable.calibration.event
questionableRegister = status.questionable.calibration.ntr
questionableRegister = status.questionable.calibration.ptr
status.questionable.calibration.enable = questionableRegister
status.questionable.calibration.ntr = questionableRegister
status.questionable.calibration.ptr = questionableRegister
The status of the questionable status calibration summary register; a zero
(0) indicates no bits set (also send 0 to clear all bits); other values indicate