System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. C / August 2016 7-363
SweepVLogMeasureI()
This KISweep factory script function performs a logarithmic voltage sweep with current measured at every step
(point).
Type TSP-Link accessible
Affected by Where saved Default value
Usage
SweepVLogMeasureI(smu, startv, stopv, stime, points)
System SourceMeter
instrument channel (for example, smua refers to SMU
Sweep start voltage in volts
Sweep stop voltage in volts
Settling time in seconds; occurs after stepping the source and before performing a
Number of sweep points (must be
≥
Details
Data for current measurements, voltage source values, and timestamps are stored in
smuX.nvbuffer1.
If all parameters are omitted when this function is called, this function is executed with the parameters
set to the default values.
Performs a logarithmic voltage sweep with current measured at every step (point):
1. Sets the smu to output startv amperes, allows the source to settle for stime seconds, and then
performs a current measurement.
2. Sets the smu to output the next volts step, allows the source to settle for stime seconds, and then
performs a current measurement.
3. Repeats the above sequence until the voltage is measured on the stopv volts step.
The source level at each step (SourceStepLevel) is automatically calculated as follows:
MeasurePoint = The step point number for a measurement
For example, for a five-point sweep (points = 5), a measurement is performed at MeasurePoint
1, 2, 3, 4, and 5.
LogStepSize = (log10(stopi) – log10(starti)) / (points – 1)
LogStep = (MeasurePoint – 1) * (LogStepSize)
SourceStepLevel = antilog(LogStep) * startv