status.measurement.reading_overflow.*
This attribute contains the measurement event reading overflow summary register set.
Type TSP-Link accessible Affected by Where saved Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B: 6
(All bits set)
Usage
measurementRegister = status.measurement.reading_overflow.condition
measurementRegister = status.measurement.reading_overflow.enable
measurementRegister = status.measurement.reading_overflow.event
measurementRegister = status.measurement.reading_overflow.ntr
measurementRegister = status.measurement.reading_overflow.ptr
status.measurement.reading_overflow.enable = measurementRegister
status.measurement.reading_overflow.ntr = measurementRegister
status.measurement.reading_overflow.ptr = measurementRegister
measurementRegister
The status of the measurement reading overflow summary register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other values indicate various
bit settings
Details
These attributes are used to read or write to the measurement event reading overflow summary
registers. Reading a status register returns a value. The binary equivalent of the returned value
indicates which register bits are set. The least significant bit of the binary number is bit B0, and the
most significant bit is bit B15. For example, assume the value 2 is returned for the enable register.
The binary equivalent is 0000 0000 0000 0010. This value indicates that bit B1 (SMUA) is set.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status register set
contents (on page E-1) and Enable and transition registers (on page E-19). The individual bits of this
register are defined in the following table.
Bit Value Description
B0
Not used
Not applicable.
B1
status.measurement.reading_overflow.SMUA
Set bit indicates that an overflow reading has
been detected for SMU A.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2
status.measurement.reading_overflow.SMUB
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set
bit indicates that an overflow reading has been
detected for SMU B.
Bit B2 decimal value: 4
Binary value: 0000 0100
B3-B15
Not used
Not applicable.