7: TSP command reference Series 2600B System SourceMeter® Instrument
7-196 2600BS-901-01 Rev. C / August 2016
smuX.abort()
This function terminates all overlapped operations on the specified source-measure unit (SMU).
Type TSP-Link accessible
Affected by Where saved Default value
Usage
smuX.abort()
Source-measure unit (SMU) channel (for example, smua.abort() applies to SMU
Details
The smuX.abort() function does not turn the output off or change any settings.
If this function is used to abort a sweep, when it is executed, the SMU exits its trigger model
immediately and returns to the trigger model's idle state.
Example
Terminates all overlapped operations on SMU channel A.
Also see
smuX.measure.overlappedY() (on page 7-225)
smuX.trigger.initiate() (on page 7-259)
smuX.buffer.getstats()
This function returns a specified reading buffer's statistics.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
statistics = smuX.buffer.getstats(bufferVar)
The statistical data about the data in the reading buffer
Source-measure unit (SMU) channel (for example, smua.buffer.getstats()
specifies SMU channel A)
The reading buffer to process
Details
This function returns a table with statistical data about the data that is placed in the buffer.
The SMU automatically updates reading buffer statistics as data is added to the reading buffer. When
the reading buffer is configured to wrap around and overwrite older data with new data, the buffer
statistics include the data that was overwritten.
The table returned from this function is a snapshot. Although the SMU continues to update the
statistics, the table returned is not updated. To get fresh statistics, call this function again.