0B System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. C / August 2016 7-317
A set bit indicates the specified timer generated an action overrun because it was still processing a
delay from a previous trigger when a new trigger was received.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status register set
contents (on page E-1) and Enable and transition registers (on page E-19). The individual bits of this
register are defined in the following table.
Bit Value Description
B1
status.operation.instrument.trigger_timer.trigger_overrun.TMR1
Bit B1 decimal
value: 2
B2
status.operation.instrument.trigger_timer.trigger_overrun.TMR2
Bit B2 decimal
value: 4
status.operation.instrument.trigger_timer.trigger_overrun.TMR3
value: 8
status.operation.instrument.trigger_timer.trigger_overrun.TMR4
status.operation.instrument.trigger_timer.trigger_overrun.TMR5
status.operation.instrument.trigger_timer.trigger_overrun.TMR6
status.operation.instrument.trigger_timer.trigger_overrun.TMR7
Bit B7 decimal
value: 128
status.operation.instrument.trigger_timer.trigger_overrun.TMR8
value: 256
B9-B15 Not used Not applicable
As an example, to set bit B1 of the operation status trigger timer trigger overrun enable register, set
status.operation.instrument.trigger_timer.trigger_overrun.enable =
status.operation.instrument.trigger_timer.trigger_overrun.TMR1.
In addition to the above constants, operationRegister can be set to the numeric equivalent of the
bit to set. To set more than one bit of the register, set operationRegister to the sum of their
decimal weights. For example, to set bits B1 and B4, set operationRegister to 18 (which is the
sum of 2 + 16).
(2
) (2
) (2
) (2
) (2
) (2
) (2
) (2
)