EasyManuals Logo
Home>Keithley>Measuring Instruments>2600B

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #398 background imageLoading...
Page #398 background image
Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-59
Details
Setting width to zero (0) seconds asserts the trigger indefinitely. To release the trigger line, use
digio.trigger[N].release().
Example
digio.trigger[4].pulsewidth = 20e-6
Sets the pulse width for trigger line 4 to
20 μs.
Also see
digio.trigger[N].assert() (on page 7-55)
digio.trigger[N].reset() (on page 7-60)
digio.trigger[N].release() (on page 7-59)
digio.trigger[N].release()
This function releases an indefinite length or latched trigger. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
digio.trigger[N].release()
N
Digital I/O trigger line (1 to 14)
Details
Releases a trigger that was asserted with an indefinite pulse width time. It also releases a trigger that
was latched in response to receiving a synchronous mode trigger. Only the specified trigger line is
affected.
Example
digio.trigger[4].release()
Releases digital I/O trigger line 4.
Also see
digio.trigger[N].assert() (on page 7-55)
digio.trigger[N].pulsewidth (on page 7-58)