Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. C / August 2016 7-173
Example
smua.reset()
smua.source.rangev = 5
smua.source.rangei = 1
smua.source.levelv = 0
smua.measure.rangev = 5
smua.measure.rangei = 1
smua.measure.nplc = 0.01
smua.measure.autozero = smua.AUTOZERO_ONCE
smua.nvbuffer1.clear()
smua.nvbuffer1.appendmode = 1
smua.source.output = smua.OUTPUT_ON
f1, msg1 = ConfigPulseVMeasureI(smua, 0, 5,
1, 0.002, 0.2, 10, smua.nvbuffer1, 1)
print(QueryPulseConfig(1).tostring())
Configure channel A to generate a pulse train,
query configuration, and then display as a string.
Channel A pulses voltage from a bias level of 0 V
to a pulse level of 5 V. The pulse level is present
for 2 ms, and the bias level for 200 ms with a 1 A
limit setting. A total of 10 pulses is generated, and
the measurement data is stored in
smua.nvbuffer1. This pulse train is assigned to
tag = 1.
Output:
>> tag = 1
>> smu = smua
>> func = volts
>> type = pulse
>> bias = 0
>> level = 5
>> limit = 1
>> time on = 0.002
>> time off = 0.2
>> points = 10
>> measure = yes
>> sync_in = 0
>> sync_out = 0
>> sync_in_timeout = 0
>> sync_out_abort = 0
>> { 5, 5, 5, 5, 5, 5, 5, 5, 5, 5 }
Also see
ConfigPulseIMeasureV() (on page 7-36)
ConfigPulseIMeasureVSweepLin() (on page 7-38)
ConfigPulseIMeasureVSweepLog() (on page 7-40)
ConfigPulseVMeasureI() (on page 7-42)
ConfigPulseVMeasureISweepLin() (on page 7-45)
ConfigPulseVMeasureISweepLog() (on page 7-47)
KIPulse factory script (on page 5-22)
reset()
This function resets commands to their default settings.
Type TSP-Link accessible
Affected by Where saved Default value
Function No
Usage
reset()
reset(system)
system
true: If the node is the master, the entire system is reset
false
: Only the local group is reset