EasyManuals Logo
Home>Keithley>Measuring Instruments>2600B

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #864 background imageLoading...
Page #864 background image
Series 2600B
System SourceMeter® Instrument Reference Manual Appendix E:
Status model
2600BS-901-01 Rev. C / August 2016 E-23
Questionable Status Registers
This register set feeds to bit B3 (QSB) of the Status Byte. The bits used in the Questionable Status
Register set are described as follows:
• Bit B8, Calibration (CAL): Set bit indicates that calibration is questionable.
• Bit B9, Unstable Output (UO): Set bit indicates that an unstable output condition was detected.
• Bit B12, Over Temperature (OTEMP): Set bit indicates that an over temperature condition was
detected.
• Bit B13, Instrument Summary (INST): Set bit indicates that a bit in the Questionable Status
Instrument Summary Register is set.
For more information on the Questionable Status Register, refer to Status register set contents
(on
page E-1) and the figures in this appendix.
Questionable Status Registers
As shown in the status model's Operation event, I/O, and TSP-Link registers (on page E-12), there
are seven register sets associated with Questionable Status. Commands are summarized in the
Status byte and service request (SRQ) (on page E-15) topic. Note that bits can also be set by using
numeric parameter values. For details, see Programming enable and transition registers (on page E-
14).
For example, either of the following commands will set the CAL enable bit (B8):
status.questionable.enable = status.questionable.CAL
status.questionable.enable = 256
When reading a register, a numeric value is returned. The binary equivalent of this value indicates
which bits in the register are set. For details, see Reading registers (on page E-15
). For example, the
following command will read the Questionable Status Enable Register:
print(status.questionable.enable)
For more information about the Questionable Status Registers, refer to Status register set contents
(on page E-1) and the figures in this appendix.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600B and is the answer not in the manual?

Keithley 2600B Specifications

General IconGeneral
BrandKeithley
Model2600B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals