System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. C / August 2016 7-349
Bit B7 B6 B5 B4 B3 B2 B1 B0
(2
) (2
) (2
) (2
) (2
) (2
) (2
) (2
)
Binary value 0/1 0/1 0/1 0/1 0/1 0/1 0/1 0/1
Example 1
enableRegister = status.system.NODE11 +
status.system.NODE14
status.system.enable = enableRegister
Uses constants to set bits B11 and B14
of the system summary enable register.
Example 2
-- decimal 18432 = binary 0100 1000 0000 0000
enableRegister = 18432
status.system.enable = enableRegister
Uses a decimal value to set bits B11 and
B14 of the system summary enable
register.
Also see
status.system2.* (on page 7-349)
System summary and standard event registers (on page E-7)
status.system2.*
These attributes manage the TSP-Link
®
system summary register of the status model for nodes 15 through 28.
These attributes are not available on the Models 2604B/2614B/2634B.
Type TSP-Link accessible Affected by Where saved Default value
.event (R) Yes Status reset Not saved 0
Usage
enableRegister = status.system2.condition
enableRegister = status.system2.enable
enableRegister = status.system2.event
enableRegister = status.system2.ntr
enableRegister = status.system2.ptr
status.system2.enable = enableRegister
status.system2.ntr = enableRegister
status.system2.ptr = enableRegister
The status of the system summary 2 register; a zero (0) indicates no bits set; other
values indicate various bit settings