EasyManua.ls Logo

Keithley 2600B - Sweepilistmeasurev()

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-358 2600BS-901-01 Rev. C / August 2016
Performs a linear current sweep with voltage measured at every step (point):
1. Sets the smu to output starti amperes, allows the source to settle for stime seconds, and then
performs a voltage measurement.
2. Sets the smu to output the next amperes step, allows the source to settle for stime seconds, and then
performs a voltage measurement.
3. Repeats the above sequence until the voltage is measured on the stopi amperes step.
The linear step size is automatically calculated as follows:
step = (stopi starti) / (points 1)
Figure 134: SweepILinMeasureV()
Example
SweepILinMeasureV(smua, -1E-3, 1E-3, 0, 100)
This function performs a 100-point linear current
sweep starting at 1 mA and stopping at +
1 mA.
Voltage is measured at every step (point) in the
sweep. Because stime is set for 0 s, voltage is
measured as quickly as possible after each
current step.
Also see
KISweep factory script (on page 5-21)
SweepIListMeasureV()
This KISweep factory script function performs a current list sweep with voltage measured at every step (point).
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
SweepIListMeasureV(smu, ilist, stime, points)
smu
System SourceMeter
®
instrument channel (for example, smua refers to SMU
channel A)
ilist
Arbitrary list of current source values; ilist = {value1, value2,
...valueN}
stime
Settling time in seconds; occurs after stepping the source and before performing a
measurement
points
Number of sweep points (must be
2)

Table of Contents

Related product manuals