7: TSP command reference Series 2600B System SourceMeter® Instrument
7-254 2600BS-901-01 Rev. C / August 2016
Details
This attribute enables or disables automatic clearing of the trigger model state machine event
detectors when the SMU transitions from the arm layer to the trigger layer.
Only the detected states of the event detectors are cleared.
The overrun statuses of the event detectors are not automatically cleared when the SMU transitions
from the arm layer to the trigger layer.
The event detectors are always cleared when a sweep is initiated.
Also see
Triggering (on page 3-32)
smuX.trigger.count
This attribute sets the trigger count in the trigger model.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes SMU reset
Instrument reset
Recall setup
Not saved 1
Usage
triggerCount = smuX.trigger.count
smuX.trigger.count = triggerCount
The trigger count is the number of times the source-measure unit (SMU) will iterate
in the trigger layer for any given sweep
SMU channel (for example, smua.trigger.count applies to SMU channel A)
Details
During a sweep, the SMU iterates through the trigger layer of the trigger model this many times. After
performing this many iterations, the SMU returns to the arm layer.
If this count is set to zero (0), the SMU stays in the trigger model indefinitely until aborted.
Also see
Triggering (on page 3-32)
smuX.trigger.endpulse.action
This attribute enables or disables pulse mode sweeps.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset